• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY EXPRESS® LD ILIAC / BILIARY; STENT, ILIAC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY EXPRESS® LD ILIAC / BILIARY; STENT, ILIAC Back to Search Results
Model Number H74938046104070
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age at time of event: probably late 70s.Device is a combination product.(b)(4).Device evaluated by mfr: the device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
It was reported that the stent dislodged and was implanted in the wrong location.The target lesion was located in the right iliac artery.A 10.0 x 40 x 75 cm express® ld iliac / biliary stent was advanced to treat the lesion; however, the stent came off the stent delivery balloon.Subsequently, a smaller balloon catheter was used to deploy the dislodged stent, followed by a larger balloon catheter to fully deploy the stent.The stent was successfully deployed but in the wrong location.The physician encountered access issues and decided to terminate the procedure.No patient complications were reported and the patient's status was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESS® LD ILIAC / BILIARY
Common Device Name
STENT, ILIAC
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7418661
Report Number2134265-2018-02946
Device Sequence Number1575348
Product Code NIO
UDI-Device Identifier08714729392095
UDI-Public08714729392095
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P090003
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 03/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/10/2020
Device Model NumberH74938046104070
Device Catalogue Number38046-10407
Device Lot Number0021243144
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2018
Initial Report FDA Received Date04/10/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-