| Model Number |
DTBA1QQ |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/19/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during initial implant the set screw was inadvertently removed from the right ventricular (rv) port on the cardiac resynchronization therapy defibrillator (crt-d).During reinsertion of the set screw, grommet damage was noticed.A new device was used and remains in use.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the device was returned and analyzed and no anomalies were found.The returned device had foreign material in the connector and on the set screw.The returned device had a damaged connector and grommet and indicated a grommet issue.The returned device indicated a set screw with a rounded socket.If information is provided in the future, a supplemental report will be issued.
|
| |
|