| Model Number |
SEDR01 |
| Medical Device Problem Code |
Pacing Problem (1439)
|
| Health Effect - Clinical Code |
Ventricular Tachycardia (2132)
|
| Date of Event |
02/22/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the physician believed that something related to the ventricular safety pacing (vsp) device feature was causing the patient to experience intermittent ventricular tachycardia (vt).The implantable pulse generator (ipg) remains is planned to be explanted and replaced as the device is also at elective replacement indicator (eri).No further patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.Product event summary: the device was returned and analyzed.Analysis of the device revealed normal battery depletion.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was further reported that the ipg was explanted and replaced.
|
| |
|