| Model Number |
YC-1800 |
| Medical Device Problem Code |
Unexpected Therapeutic Results (1631)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The ophthalmologist observed pitting on the posterior surface of the lens while performing surgery.However, the ophthalmologist completed the surgical procedure by performing certain modifications to achieve the intended results.Nidek inc.Has learned that dr.(b)(6) has doubts the effectiveness/reliability of the device in question and therefore, nidek has decided to replace it with a new yc-1800.The device is not yet evaluated by nidek inc.A follow-up report to this mdr will be submitted upon the completion of device evaluation.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the customer contacted nidek inc.Field service manager who later communicated to nidek inc.Customer service manager to report that the doctor observed pitting lens during surgical procedure.A total of 3 incidents were reported and a separate mdr will be submitted for each case.Two patients have acrysoft sn60wf monofocal lens and one have multifocal lens.There was no serious adverse event reported from the user facility but nidek inc.Considers pitting lens issue on yc-1800 a reportable event as it is an undesirable condition and has a potential to cause or contribute to a serious injury if the issue were to recur.
|
| |
|
Additional Manufacturer Narrative
|
|
Service on the device yc-1800 was performed at the user facility by a nidek inc.Field service engineer (fse) on 01-may-2018.As a result of the investigation by the fse, the reported failure could not be confirmed.Fse powered on the yc-1800 s/n y1620072 and noticed the shot counter was 03x1000.Fse did not evaluate the system and it was returned to nidek inc.For evaluation.A new yc-1800 s/n y1620091 was installed at the user facility.Fse verified shot counter = 01 x 1000 and the following parameters: - proper focus and alignment of the yag laser beam / aiming beam.- offset focus shift is proper 50 -1500 microns, anterior and posterior - energy output is spec, 0.2 - 12.0mj single, 0.4 - 22.2mj double burst, 0.6 - 31.0mj triple burst - optical breakdown at 3.3mj - tested and inspected the new system.System is operational.The yc-1800 s/n (b)(4).Was evaluated by a nidek inc.Servide engineer (se) on (b)(6)2018 but could not duplicate the customer's reported problem.Se found plasma discharge at 6.2 mj.Standard plasma discharge should be 5mj or less.Se replaced laser head assembly and plasma discharge at 3.8mj is 100% after replacement.Se calibrated pfn (pulsed form network) voltage = 390vdc and focus and alignment of yag/aiming beam.Laser output was verified to be 0.2mj-11.5mj in spec.The system is operational at all the settings.Should new information that changes the facts and/or conclusion of this report become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
Please refer to initial mdr for details.
|
| |
|
Search Alerts/Recalls
|