| Medical Device Problem Codes |
Moisture Damage (1405); Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a casing/condition (case damage with moisture) issue.It was alleged that the battery compartment was damaged with moisture in it.There was no indication that the product caused or contributed to an adverse event.This complaint is being reported because the alleged malfunction has the ability to result in a delay in treatment or long term cessation in delivery if the damage impacts the power circuit or cartridge compartment.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1: date of submission (b)(6) 2018 device evaluation: the pump has been returned and evaluated by product analysis on (b)(6) 2018 with the following findings: during investigation, moisture was found in the battery compartment.A leak test failed due to cracks in the battery compartment along the side and from the case seal traveling to the bumper.The pump was opened to find no moisture.
|
| |
|