According to the reporter, intra-operatively, the two device¿s blade snapped off four times during activation.It was suspected to be in contact with metal.The broken pieces that fell into patients cavity are all retrieved.Another device was used to complete the procedure.There was no patient injury.
|
|
Evaluation summary: two devices were returned for evaluation.These devices have been used in the treatment or diagnosis of a patient.A review of the lot number reported indicates that the product was within the assigned expiration date at the time of the reported incident.The returned products did not meet specification as received.Visual inspection of each of the two disposable devices revealed that the tip of the waveguide was broken off.The broken pieces were returned.Investigation personnel performed functional testing on the returned hand piece using a test lab generator and battery.For each device, the assembled device returned a green light and a welcome tone, but immediately returned a red light-emitting diode (led) indicator with an error tone (alarm activation) when the device was activated.This characteristic indicated that the device was not functional.The waveguide had its tip missing, and the missing piece was returned.Investigation personnel concluded that the titanium waveguide was in use when it cracked and broke off, and may have come in contact with any of the following; hemostats, clips, staples, retractors, etc.During use.The investigation identified the cause of the reported event to be user error.The device instructions for use currently contain a warning against contact between the dissector tip and metal objects (hemostats, clips, staples, retractors, etc.) during activation.Contact with metal objects during use will cause the active blade to crack and may eventually break off.This issue is specific to ultrasonic dissectors.The instructions for use (ifu) advises device users to visually inspect all system components for breaks, cracks, nicks, or other damages prior to use.Instructions for use (ifu) also states: do not use damaged components.Use of damaged components may result in injury to the patient or user.No update to the risk management file is required at this time.1717344-2018-00470-1 if information is provided in the future, a supplemental report will be issued.
|