| Model Number |
M0068403931 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/27/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Investigation results: one flexiva 550 laser fiber was received for evaluation.Visual examination revealed that the exposed glass fiber tip measured approximately 3.5 mm and appeared unused.Additional examination under magnification confirmed that the tip was unused.There was no damage along the body of the fiber.The investigator was unable to examine the fiber face within the sub miniature-a (sma) connector because light was unable to travel to the fiber face, this indicated that the fiber was broken within the connector.The condition of the returned device would cause a weak aiming beam thereby confirming the event.Damage was caused to the device without direct patient contact.Therefore, a review and analysis of all available information indicated that the most probable root cause is handling damage.A review of the device history record (dhr) was performed; no anomalies were noted.A search of the complaint database revealed that no similar complaints exist for the specified lot.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that a flexiva 550 laser fiber was used during a ureteroscopy procedure in the ureter performed on (b)(6) 2018 according to the complainant, during preparation and outside the patient, after the laser fiber was connected into the laser console, the aiming beam was not visible.The procedure was completed with another flexiva 550 laser fiber.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be fine.Note: this event has been deemed an mdr-reportable event based on investigation results which revealed that the laser fiber was broken within connector.
|
| |
|
Search Alerts/Recalls
|