Device evaluated by mfr: visual inspection of the returned device did not reveal any defects.The steering knob and tension control knob functioned properly on both the lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.Both right and left curves failed dimensional testing.The shaft was measured in three sections and was within specification.X-ray revealed multiple collapsed guide coils near the distal end of the device, which hindered movement of the steering wires to achieve the proper bend at the distal end.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion for the guide coil collapse is design specification as the device problem was traced back to the design specifications.Updated: device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes.(b)(4).
|