Model Number M00420SL28600
Date of Event 03/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported the catheter became entangled in the patient¿s anatomy.A blazer¿ dx ablation catheter was selected for use in an atrial fibrillation ablation procedure.The catheter was introduced into the femoral vein and was being pushed up into the heart when it prolapsed on itself and then became entangled in the tricuspid valve.The physician worked to untangle the catheter from the tissue and the steering mechanism stopped working.The physician introduced a snare to dislodge the catheter from the tricuspid valve.The catheter was removed and another manufacturer¿s catheter was selected to complete the procedure.There were no patient complications.
 
Additional Manufacturer Narrative
Device evaluated by mfr: visual inspection of the returned device did not reveal any defects.The steering knob and tension control knob functioned properly on both the lock and unlock positions.No abnormal resistance was felt when actuating the steering mechanism.Both right and left curves failed dimensional testing.The shaft was measured in three sections and was within specification.X-ray revealed multiple collapsed guide coils near the distal end of the device, which hindered movement of the steering wires to achieve the proper bend at the distal end.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion for the guide coil collapse is design specification as the device problem was traced back to the design specifications.Updated: device evaluated by mfr., eval summary attached, method codes, result codes, conclusion codes.(b)(4).
 
Event or Problem Description
It was reported the catheter became entangled in the patient¿s anatomy.A blazer¿ dx ablation catheter was selected for use in an atrial fibrillation ablation procedure.The catheter was introduced into the femoral vein and was being pushed up into the heart when it prolapsed on itself and then became entangled in the tricuspid valve.The physician worked to untangle the catheter from the tissue and the steering mechanism stopped working.The physician introduced a snare to dislodge the catheter from the tricuspid valve.The catheter was removed and another manufacturer¿s catheter was selected to complete the procedure.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLAZER¿ DX-20 CATHETER
Common Device Name
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
MDR Report Key7421251
Report Number2134265-2018-03298
Device Sequence Number61755
Product Code DRF
UDI-Device Identifier08714729780533
UDI-Public08714729780533
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K081576
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/22/2018
1 Device was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/02/2021
Device Model NumberM00420SL28600
Device Catalogue Number20SL2860
Device Lot Number0021562643
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2018
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer05/07/2018
Initial Report FDA Received Date04/11/2018
Supplement Report FDA Received Date05/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
-
-