• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910  ZT TRIAL SLEEVE 22D SML; HIP INSTRUMENTS : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910  ZT TRIAL SLEEVE 22D SML; HIP INSTRUMENTS : FEMORAL TRIALS Back to Search Results
Catalog Number 257622110
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that identifying marks found to be damaged and the pin was bent in post cleaning inspection.
 
Additional Manufacturer Narrative
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZT TRIAL SLEEVE 22D SML
Common Device Name
HIP INSTRUMENTS : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key7421521
Report Number1818910-2018-57132
Device Sequence Number10775372
Product Code LXH
UDI-Device Identifier10603295146902
UDI-Public10603295146902
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number257622110
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2018
Initial Date Received by Manufacturer 03/14/2018
Supplement Date Received by Manufacturer04/13/2018
07/26/2018
Initial Report FDA Received Date04/11/2018
Supplement Report FDA Received Date04/24/2018
08/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-