Intuitive surgical inc.(isi) received the hrsv monitor involved with this complaint and completed the device evaluation.Failure analysis found that the hrsv monitor does not turn on.The unit power and ic caps were repaired/replaced.Also, the unit was cleaned, calibrated and tested.The complaint is being reported due to a da vinci system malfunction rendering the da vinci system unavailable for use after the start of a surgical procedure.Although no patient harm occurred, if the reported malfunction were to recur, it could cause or contribute to an adverse event.
|
It was reported that during a da vinci-assisted surgical procedure, the left eye was missing from the surgeon side console (ssc).The intuitive surgical inc.(isi) technical support engineer (tse) instructed the site to perform a restart and hard restart of the system but the issue could not be resolved.The surgeon decided to use another ssc from another system to complete the case.There was no report of patient harm, adverse outcome or injury.An isi field service engineer (fse) was dispatched to the site to further investigate the reported complaint.The fse confirmed the reported issue and replaced left high resolution stereo viewer (hrsv) monitor.The hrsv is located on the surgeon side console and provides the 3d image captured from the endoscope inserted in the patient.
|