Catalog Number 0204511
Medical Device Problem Codes Retraction Problem (1536); Material Rupture (1546); Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was not provided, a manufacturing review could not be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during a tavr procedure, the balloon allegedly ruptured during an inflation attempt.It was further reported that there was difficulty retracting the balloon through the 14fr sheath.Upon removing the balloon and sheath together as one unit, a segment of the balloon material allegedly detached and became lodged in the external iliac artery.Multiple attempts to remove the balloon segment including upsizing to a 16fr sheath and the use of a snare were unsuccessful.The health care provider decided to send the patient to surgery for removal of the balloon segment.There was no reported patient injury post surgical extraction of the balloon segment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUE DILATATION CATHETER
Common Device Name
BALLOON VALVULOPLASTY CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7421972
Report Number2020394-2018-00367
Device Sequence Number69719
Product Code OZT
UDI-Device Identifier00801741090998
UDI-Public(01)00801741090998
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K150667
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other,use
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0204511
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2018
Initial Report FDA Received Date04/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age82 YR
Patient Weight87
-
-