• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC CRESCENTIC BLADE 24MM RADIUS/45MMX0.6MM; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC CRESCENTIC BLADE 24MM RADIUS/45MMX0.6MM; BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 03.000.394
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/01/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
This is report 1 of 3 for the same event: it was reported that the three crescentic saw blade devices always lost their teeth on the same side.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Additional Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and the reported condition was confirmed.A visual and functional assessment was performed which found that the left most teeth of the device were broken off.It was further determined that the cause of the crescentic saw blades teeth to break was due to the design.This issue has been escalated to capa.If information is obtained that was not available, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRESCENTIC BLADE 24MM RADIUS/45MMX0.6MM
Common Device Name
BLADE, SAW, GENERAL AND PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key7422367
Report Number8030965-2018-52989
Device Sequence Number39257
Product Code GFA
UDI-Device Identifier07611819996021
UDI-Public(01)07611819996021(11)140516(10)MA1011
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number03.000.394
Device Lot NumberMA1011
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2018
Initial Date Received by Manufacturer 03/14/2018
Supplement Date Received by Manufacturer04/09/2018
05/16/2018
Initial Report FDA Received Date04/11/2018
Supplement Report FDA Received Date04/17/2018
05/23/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-