• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125225-15
Health Effect - Clinical Code Vessel Or Plaque, Device Embedded In (1204)
Date of Event 03/20/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.The investigation determined the reported difficulties and treatment appear to be related to circumstances of the procedure as it is likely the device interacted with the difficult anatomy causing the reported failure to advance and subsequent stent dislodgement.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion with heavy tortuosity and heavy calcification in the obtuse marginal coronary artery after the patient had refused to have coronary artery bypass (cabg) surgery.A 2.25 x 15 mm xience alpine stent delivery system (sds) was advanced in the vessel, but met resistance due to the circumflex having an acute angulation and being very tortuous.When the sds was withdrawn from the patient anatomy, it was noted the stent had dislodged in a calcium bed in the vessel.A snare device was used to retrieve the dislodged stent but was unsuccessful.A 3.5 x 15 mm xience alpine stent was deployed crushing the dislodged stent against the calcified artery wall.Therefore, the procedure was aborted as the patient had severe coronary artery disease and had refused cabg.The patient will continue with medical management.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7422555
Report Number2024168-2018-02658
Device Sequence Number39323
Product Code NIQ
UDI-Device Identifier08717648199073
UDI-Public08717648199073
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/13/2021
Device Catalogue Number1125225-15
Device Lot Number8020641
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2018
Initial Report FDA Received Date04/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient Weight90
-
-