The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction centrifuge bowl leak/break.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot f901 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot f901 for the reported issue shows no trends.Trends were reviewed for complaint categories, centrifuge bowl leak/break and leak centrifuge alarm no trends were detected for each complaint category.A photograph analysis was conducted for this complaint.A review of the customer provided photographs show the leak centrifuge alarm on the monitor screen.The additional photographs provided show the centrifuge bowl installed onto the instrument, with no signs of a blood or liquid leak on any external surface.The origin of the leak could not be determined from the photographs.A material trace on the bowls used to build lot f901 found no related nonconformances.The device history record review did not result in any related non-conformances and this kit lot had passed all lot release testing.The root cause of the centrifuge bowl leak/break could not be determined based on the information provided.This investigation is now complete.(b)(4).
|