• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IMMUCOR, INC. CAPTURE-R READY-ID; REAGENT RED BLOOD CELLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IMMUCOR, INC. CAPTURE-R READY-ID; REAGENT RED BLOOD CELLS Back to Search Results
Catalog Number 0066204
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Immucor technical support used a remote electronic connection method on (b)(6) 2018 to assess test well images in question, which showed that they were visually negative.The immucor laboratory was unable to test id353 retention product because the product had already expired before the investigation.Therefore immucor performed a dhr review of id353 which indicated that the product performed as expected at the time of release, and that all specifications were deemed to be acceptable before the id353 product was released to market.
 
Event or Problem Description
On (b)(6) 2018, a customer site reported an unexpected negative antibody identification when using capture-r ready-id when used on a galileo echo instrument, when tested on (b)(6) 2018.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURE-R READY-ID
Common Device Name
REAGENT RED BLOOD CELLS
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 5625
MDR Report Key7422589
Report Number1034569-2018-00092
Device Sequence Number10780681
Product Code KSZ
UDI-Device Identifier10888234001140
UDI-Public10888234001140
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
102707/0.0
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Medical Technologist
Type of Report Initial
Report Date (Section B) 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Medical Technologist
Device Expiration Date03/20/2018
Device Catalogue Number0066204
Device Lot NumberID353
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2018
Initial Report FDA Received Date04/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age83 YR
-
-