• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDELA LLC BREASTPUMP PNSA STARTER; PUMP, BREAST, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDELA LLC BREASTPUMP PNSA STARTER; PUMP, BREAST, POWERED Back to Search Results
Model Number 57081
Health Effect - Clinical Code Inflammation (1932)
Date of Event 02/10/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
In follow up with the customer related to her mastitis allegation with the swing pump, on (b)(6) 2018 she indicated that she experienced mastitis, which started on (b)(6) 2018, and then recurred on (b)(6) 2018, while using both the pump in style and swing breast pumps, as well as the replacement pump in style which was sent to her in response to her original report on (b)(6) 2018.She indicated that she was prescribed various courses of antibiotics and she just finished the last course and was hopeful that the mastitis would resolve.She also indicated that even after trying the swing pump with the power supply and new batteries, she continued to have issues, so a replacement swing pump was sent to the customer and return of her original swing pump was requested.In further follow up with the customer on (b)(6) 2018, she indicated that she was hospitalized for the mastitis and was being treated with iv antibiotics.In the hospital she was using a symphony breast pump and found that it was effective, so when she is released from the hospital, she is going to look into renting one.This medwatch report 1419937-2018-00101 relates to the original pump in style, medwatch report 1419937-2018-00102 relates to the swing and medwatch report 1419937-2018-00103 relates to the replacement pump in style.The original pump in style device was returned without the customer's parts and accessories; therefore, it was evaluated with a medela lab kit on (b)(6) 2018 and it passed suction and cycle specifications and did not generate a grinding noise, even though it was noted during the evaluation that there was damage to the retention snaps and the back plate mounting posts.Refer to attached evaluation.The customer's report of low suction related to the original pump in style could not be confirmed.Based on the results of (b)(4), it cannot be definitively concluded that the pump caused or contributed to the customer¿s mastitis.The estimated incidence of mastitis in lactating women, whether using a breast pump or not, according to published clinical literature can be as high as (b)(6)%.In fact, clinical guidelines suggest the use of a breast pump to facilitate withdrawal of breast milk during bouts of mastitis.The complaint rate of mastitis across all reported failures, across all medela breast pumps, is (b)(4)% for the period of (b)(6) 2013 to (b)(6) 2017.Mastitis is usually a benign, self-limiting infection with few consequences for the suckling infant.The risk of mastitis is higher among women who have breastfed previously, especially those with a history or mastitis." riordan & wambach, 4th ed.P.294: breastfeeding and human lactation.Mastitis requires prompt medical attention for the mother for pain relief and prescription antibiotics to avoid progression to overwhelming sepsis.(b)(4).
 
Event or Problem Description
On (b)(6) 2018, the customer alleged to medela llc via email that the pump in style breast pump which she received through her insurance, which she had been using for 5 weeks, was making a strange "grinding sound." she additionally alleged that she rarely double pumps with it because the suction "suffers." at this time, the customer made no mention of mastitis.Customer service emailed the customer troubleshooting tips, which helped her to identify an issue with the faceplate mounting posts, which she reported on (b)(6) 2018.At that time, the customer was sent a replacement device.On (b)(6) 2018, the same customer alleged to medela llc via email that she purchased a swing breast pump to use in addition to the pump in style and it was "terribly ineffective." she alleged that after using it for 10 days, the suction was negligible even at high levels and that she gets little milk compared to what she gets when using the pump in style.She further alleged that she had chronic mastitis.Customer service encouraged the customer to phone in so that her issue could be addressed.She phoned in on (b)(6) 2018 and indicated that she would test the pump using the power supply and also with new batteries.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BREASTPUMP PNSA STARTER
Common Device Name
PUMP, BREAST, POWERED
Manufacturer (Section D)
MEDELA LLC
1101 corporate dr
mchenry IL 60050
MDR Report Key7422633
Report Number1419937-2018-00101
Device Sequence Number39292
Product Code HGX
UDI-Device Identifier00020451570815
UDI-Public0100020451570815
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K031614
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial
Report Date (Section B) 04/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model Number57081
Device Catalogue Number57081
Device Lot Number44036
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2018
Distributor Facility Aware Date04/06/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2018
Initial Report FDA Received Date04/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-