Event: additional information evaluation / investigation: the actual complaint device was not returned.Without the complaint device, a physical investigation was not able to be completed.No imaging was provided for review.A review of the device history record, instructions for use, manufacturing instructions and quality control data was conducted.The device history record was reviewed and no non-conformities were noted.Thrombosis can occur for a variety of reasons, such as genetic predisposition, iliac tortuosity, external compression, stenosis, improper overlap, narrow inner iliac lumen, vessel damage, and rough surfaces.Tortuosity, compression, and pulsation can create high shear forces that stimulate thrombus growth.Where a small thrombus has formed, the clot can enlarge because the blood flow slows around the clot, allowing clot-forming elements to be deposited.Pulsation and a small inner lumen can occur through graft oversizing or undersizing.Patient non-compliance could have also led to this occurrence.In the instructions for use (ifu) provided with the device, it is stated that "all patients should be monitored closely and checked periodically for a change in the condition of their disease and the integrity of the endoprosthesis." since it was reported that the patient was lost to follow up, the thrombus formation could not be detected until the patient presented to the physician with claudication.This likely increased the severity of this event.The physician reported that there was no tortuosity or calcification that was of significance.Planning and sizing information and imaging of the event was requested but not provided.The explanted devices were requested but not provided.This information would have aided in determining a definitive cause for this event.After review of the design history files and device history record, there is no evidence to suggest that the device was manufactured out of specifications.Genetic predisposition, improper sizing, improper device placement, patient non-compliance, and disease progression cannot be ruled out as possible causes for this complaint.A definitive cause cannot be established with the information available.Measures have been initiated to address this failure mode.Monitoring will continue to be performed for similar complaints.The appropriate cook personnel have been notified of this event.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Evaluation / investigation: the actual complaint device was not returned.Without the complaint device, a physical investigation was not able to be completed.No imaging was provided for review.A review of the device history record, instructions for use, manufacturing instructions and quality control data was conducted.The device history record was reviewed and no non-conformities were noted.Thrombosis can occur for a variety of reasons, such as genetic predisposition, iliac tortuosity, external compression, stenosis, improper overlap, narrow inner iliac lumen, vessel damage, and rough surfaces.Tortuosity, compression, and pulsation can create high shear forces that stimulate thrombus growth.Where a small thrombus has formed, the clot can enlarge because the blood flow slows around the clot, allowing clot-forming elements to be deposited.Pulsation and a small inner lumen can occur through graft oversizing or undersizing.Patient non-compliance could have also led to this occurrence.In the instructions for use (ifu) provided with the device, it is stated that "all patients should be monitored closely and checked periodically for a change in the condition of their disease and the integrity of the endoprosthesis." since it was reported that the patient was lost to follow up, the thrombus formation could not be detected until the patient presented to the physician with claudication.This likely increased the severity of this event.The physician reported that there was no tortuosity or calcification that was of significance.Planning and sizing information and imaging of the event was requested but not provided.The explanted devices were requested but not provided.This information would have aided in determining a definitive cause for this event.After review of the design history files and device history record, there is no evidence to suggest that the device was manufactured out of specifications.Genetic predisposition, improper sizing, improper device placement, patient non-compliance, and disease progression cannot be ruled out as possible causes for this complaint.A definitive cause cannot be established with the information available.Measures have been initiated to address this failure mode.Monitoring will continue to be performed for similar complaints.The appropriate cook personnel have been notified of this event.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|