A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.Unit passed their dat test with 0.08795 inches.All operating currents are within specification.No unexpected low battery or off no power alarms noted.Unit passed functional test including displacement test, rewind, basic occlusion, occlusion, prime/compromised force sensor system, and excessive no delivery alarm test.Bolus history and daily totals appear on the main menu screen properly.Unit functioned properly during testing.Unit was received with minor scratched lcd window, cracked reservoir tube lip and cracked battery tube threads.
|