• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551LNAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551LNAS; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551LNAS
Device Problems Break (1069); Inaccurate Delivery (2339)
Patient Problem Hypoglycemia (1912)
Event Date 01/21/2018
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
Customer called and reported that they experienced severe lows and received emergency medical assistance due to low blood glucose on (b)(6) 2018 with blood glucose of 40 mg/dl at the time of the incident.The customer also had lows on (b)(6) and (b)(6) 2018.The customer was given glucagon shots by the school nurse to treat.The customer was most likely wearing the insulin pump during the incident.The customer stated that the pump was broken and not displaying history, so their health care professional took them off the pump.Troubleshooting was not completed as the customer declined.The customer is currently using manual injections to treat their blood glucose.The insulin pump will not be returned for analysis.
 
Manufacturer Narrative
A complete analysis and testing of the insulin pump showed that it was functioning properly and passed all functional testing.After testing it was concluded that the device operated within specifications.Unit passed their dat test with 0.08795 inches.All operating currents are within specification.No unexpected low battery or off no power alarms noted.Unit passed functional test including displacement test, rewind, basic occlusion, occlusion, prime/compromised force sensor system, and excessive no delivery alarm test.Bolus history and daily totals appear on the main menu screen properly.Unit functioned properly during testing.Unit was received with minor scratched lcd window, cracked reservoir tube lip and cracked battery tube threads.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551LNAS
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
gerwin de graaff
ceiba norte ind. park #50 road
juncos 00777--386
MDR Report Key7423502
MDR Text Key105239902
Report Number3004209178-2018-63977
Device Sequence Number1
Product Code OZO
UDI-Device Identifier00643169507739
UDI-Public(01)00643169507739
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 04/30/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-551LNAS
Device Catalogue NumberMMT-551LNAS
Device Lot NumberA4551LNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received04/17/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age42 YR
Patient Weight182
-
-