| Model Number |
M006175253180 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a percuflex¿ plus stent was used for a procedure performed on (b)(6) 2018.According to the complainant, during unpacking it was noticed that the stent was broken into two pieces.The box and bag that the stent arrived in was noted to have no visible damage.The procedure was completed with another stent of a different size.Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date.Should additional relevant details become available; a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
A percuflex plus stent was returned for analysis.A visual analysis of the returned device found that the bladder pigtail was detached.No other anomalies or damages were noted.The complaint was confirmed.Based on product analysis, the failure was noted to have occurred outside the patient and t during unpacking.It is possible that the device was handled and manipulated in a way that contributed to the failure encountered -stent detached, and the defect could have been caused by the user due to excessive manipulation and interaction of the stent with suture.This type of failure is consistent with one caused when a device is pulled strongly and was submitted to an excess of force; excess of force may cause damage, deformities or device break.Therefore, the most probable root cause of this complaint is handling damage.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications at the time of release for distribution.A search of the complaint database revealed that no similar complaints exist for the specified lot.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation that a percuflex plus stent was used for a procedure performed on (b)(6) 2018.According to the complainant, during unpacking it was noticed that the stent was broken into two pieces.The box and bag that the stent arrived in was noted to have no visible damage.The procedure was completed with another stent of a different size.Attempts to obtain additional information regarding the circumstances surrounding this event have been unsuccessful to date.Should additional relevant details become available; a supplemental report will be submitted.
|
| |
|