• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARSTREAM TECHNOLOGIES LTD. LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT Back to Search Results
Model Number LSMU1350526
Medical Device Problem Code Occlusion Within Device (1423)
Health Effect - Clinical Code Stenosis (2263)
Date of Event 02/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Manufacturing review: the lot number has been provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported for this lot number and issue to date.Investigation summary: the result of the investigation is inconclusive.The sample was not returned for evaluation.Based upon the available information a definitive root cause has not been determined.Labeling review: the ifu for the lifestream product was reviewed and contains the following information relevant to the reported event: a device description: 1.Implant the lifestream¿ balloon expandable vascular covered stent is comprised of an electropolished balloon-expandable stent made from (b)(4) stainless steel, encapsulated between two layers of eptfe.B indication for use: the lifestream¿ balloon expandable vascular covered stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries.Directions for use: site access and preparation: 1.Using standard techniques access the artery and place an introducer sheath or guiding catheter of appropriate inner diameter and a 0.035" (0.89 mm) guidewire across the target lesion.2.Perform diagnostic angiography to confirm site of implantation and measure the reference vessel diameter and lesion length.Covered stent size selection: 3.Select a covered stent diameter that is approximately 5%-20% larger than the largest reference vessel diameter at the proximal or distal target site.Refer to the sizing table on the packaging label for appropriate selection of the covered stent diameter and length.Endovascular system preparation: 4.Carefully remove the selected device from the package.5.Inspect the covered stent for adherence to the balloon and centered placement in relation to the balloon marker bands.If the covered stent is not centered and/or does not firmly adhere to the balloon, do not use.6.Flush the delivery system guidewire lumen with sterile saline mixture until saline drops from the distal end of the endovascular system.Air evacuation: 7.A 20 cc or smaller luer-lock syringe with a minimum of 5 cc¿s sterile saline mixture is recommended for use for aspirating this device.8.With the distal balloon tip pointing down and positioned below the level of the syringe, pull negative pressure until all air is expelled.9.Induce a negative pressure to remove any air from the balloon and inflation lumen.Repeat until all air is expelled.10.Carefully release to neutral.Allow the inflation lumen to fill with the diluted contrast medium and maintain a neutral pressure.Important: do not apply positive pressure to the balloon.11.Attach the prefilled inflation device to the inflation lumen of the catheter hub, ensuring no air bubbles remain at the catheter connection.12.Verify that the covered stent is still centered between the two radiopaque markers on the balloon catheter.Introduction of the endovascular system and placement of the covered stent: 13.Advance the endovascular system over the guidewire into the introducer sheath.14.Further advance the endovascular system to the target treatment site within the introducer sheath and position the covered stent across the lesion.Verify that the covered stent is still centered within the balloon marker bands.Slowly retract the introducer sheath / guiding catheter while maintaining the position of the covered stent.Ensure the introducer sheath is retracted far enough to not compromise the balloon expansion and covered stent release.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported through the results of a clinical trial that approximately thirty-six months post stent graft placement in the right external iliac artery, duplex ultrasound demonstrated stenosis just proximal to the proximal edge of the stent.There was no reported intervention.The current patient status was not provided.
 
Additional Manufacturer Narrative
H10: after further clinical review, this event was reassessed and determined to be not mdr reportable.The stenosis was identified to be proximal to the proximal edge of the study device and did not involve the study device.Although, this event is not mdr reportable, an initial emdr has already been submitted; therefore, this supplemental report is being submitted to document the change in reportability and any new or additional complaint details.Manufacturing review: the lot number has been provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.This is the only complaint reported for this lot number and issue to date.Investigation summary: the result of the investigation is inconclusive.The sample was not returned for evaluation.It is understood that the complaint stent was not responsible for the stenosis an it was confirmed that the stenosis location was proximal to the stent and did not involve the study device.Therefore it is likely that others factors possible patient factors contributed to the reported stenosis.Labeling review: the ifu for the lifestream product was reviewed and contains the following information relevant to the reported event: a device description 1.Implant the lifestream¿ balloon expandable vascular covered stent is comprised of an electropolished balloon-expandable stent made from 316l stainless steel, encapsulated between two layers of eptfe.B indication for use the lifestream¿ balloon expandable vascular covered stent is indicated for the treatment of atherosclerotic lesions in common and external iliac arteries.Directions for use site access and preparation 1.Using standard techniques access the artery and place an introducer sheath or guiding catheter of appropriate inner diameter and a 0.035" (0.89 mm) guidewire across the target lesion.2.Perform diagnostic angiography to confirm site of implantation and measure the reference vessel diameter and lesion length.Covered stent size selection 3.Select a covered stent diameter that is approximately 5%-20% larger than the largest reference vessel diameter at the proximal or distal target site.Refer to the sizing table on the packaging label for appropriate selection of the covered stent diameter and length.Endovascular system preparation 4.Carefully remove the selected device from the package.5.Inspect the covered stent for adherence to the balloon and centered placement in relation to the balloon marker bands.If the covered stent is not centered and/or does not firmly adhere to the balloon, do not use.6.Flush the delivery system guidewire lumen with sterile saline mixture until saline drops from the distal end of the endovascular system.Air evacuation 7.A 20 cc or smaller luer-lock syringe with a minimum of 5 cc¿s sterile saline mixture is recommended for use for aspirating this device.8.With the distal balloon tip pointing down and positioned below the level of the syringe, pull negative pressure until all air is expelled.9.Induce a negative pressure to remove any air from the balloon and inflation lumen.Repeat until all air is expelled.10.Carefully release to neutral.Allow the inflation lumen to fill with the diluted contrast medium and maintain a neutral pressure.Important: do not apply positive pressure to the balloon.11.Attach the prefilled inflation device to the inflation lumen of the catheter hub, ensuring no air bubbles remain at the catheter connection.12.Verify that the covered stent is still centered between the two radiopaque markers on the balloon catheter.Introduction of the endovascular system and placement of the covered stent 13.Advance the endovascular system over the guidewire into the introducer sheath.14.Further advance the endovascular system to the target treatment site within the introducer sheath and position the covered stent across the lesion.Verify that the covered stent is still centered within the balloon marker bands.Slowly retract the introducer sheath / guiding catheter while maintaining the position of the covered stent.Ensure the introducer sheath is retracted far enough to not compromise the balloon expansion and covered stent release.
 
Event or Problem Description
It was reported through the results of a clinical trial that approximately thirty-six months post stent graft placement in the right external iliac artery, duplex ultrasound demonstrated stenosis just proximal to the proximal edge of the stent.There was no reported intervention.The current patient status was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTREAM BALLOON EXPANDABLE VASCULAR COVERED STENT
Common Device Name
BALLOON EXPANDABLE VASCULAR COVERED STENT
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford
EI 
MDR Report Key7423866
Report Number9616666-2018-00042
Device Sequence Number16417513
Product Code PRL
UDI-Device Identifier05391522080980
UDI-Public(01)05391522080980
Combination Product (Y/N)N
Initial Reporter CountryNZ
PMA/510(K) Number
P160024
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2015
Device Model NumberLSMU1350526
Device Catalogue NumberLSMU1350526
Device Lot Number50111803
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/19/2018
Supplement Date Received by Manufacturer04/30/2018
Initial Report FDA Received Date04/11/2018
Supplement Report FDA Received Date05/22/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ASPIRIN, CLOPIDOGREL; ASPIRIN, CLOPIDOGREL
Patient Age71 YR
-
-