WRIGHTS LANE SYNTHES USA PRODUCTS LLC ROD INTRODUCTION PLIERS FOR DUAL-OPENING IMPL F/6.0MM RODS; PLIERS,SURGICAL Back to Search Results
Model Number 388.509
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device used for treatment, not diagnosis.Patient information not available for reporting.Device is an instrument and is not implanted/explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Device history records review not available to report at this time.The investigation could not be completed; no conclusion could be drawn, as no product was received.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that one (1) table top rod cutter and bender had a lot of resistance, causing a rod to not pass in and out through the holes in the cutter, and one (1) rod introduction pliers for dual-opening implants for 6.0 mm rods caused the scrub technicians a hard time loading an implant on it during a spinal fusion procedure on (b)(6) 2018, and subsequently, the handle on the pliers fell off at an unknown time.The scrub technicians were training a new recruit during a surgery when they loaded an unknown collar on the pliers and the pliers did not hold the collar as intended, giving the scrub technicians a hard time.Subsequently, the handle on the pliers was found to have fallen off after cleaning in the sterile processing department (spd) and it is unknown if the handle fell off during the surgery or during the cleaning process.Another set of pliers and cutter/bender were available at the time of the surgery and hence there was no surgical delay.The procedure was reported to have been completed successfully and without any complications.The patient condition was reported to have been stable.This complaint involves one (1) instruments.Concomitant devices reported: unknown collar (part# unknown, lot# unknown, quantity# 1); unknown 6.0 mm rods (part# unknown, lot# unknown, quantity# unknown).This report is 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Device history records review was completed for part# 388.509, lot# a7na31.Manufactured by: (b)(4), manufacturing date: aug 26, 2004.No non- conformance reports were generated during production.Review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Service and repair evaluation was completed.The customer reported the handle of the pliers fell off after it was difficult to load the implant.The repair technician reported the spring and ratchet were worn.Worn out parts is the reason for repair.The cause of the issue is unknown.The following parts were replaced: spring (set), ratchet.The item was repaired per the inspection sheet, passed synthes final inspection on 6-apr-2018 and will be returned to the customer upon completion of the service and repair process.The device was deemed serviceable and returned to the customer, no design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Device history records review was completed for part# 388.509, lot# a7na31.Manufactured by: (b)(4), manufacturing date: aug 26, 2004.No non- conformance reports were generated during production.Review of the device history records showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Service and repair evaluation was completed.The customer reported the handle of the pliers fell off after it was difficult to load the implant.The repair technician reported the spring and ratchet were worn.Worn out parts is the reason for repair.The cause of the issue is unknown.The following parts were replaced: spring (set), ratchet.The item was repaired per the inspection sheet, passed synthes final inspection on 6-apr-2018 and will be returned to the customer upon completion of the service and repair process.The device was deemed serviceable and returned to the customer, no design or manufacturing issues were identified therefore it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROD INTRODUCTION PLIERS FOR DUAL-OPENING IMPL F/6.0MM RODS
Common Device Name
PLIERS,SURGICAL
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7424085
Report Number2939274-2018-51604
Device Sequence Number112577
Product Code HTC
UDI-Device Identifier10705034773339
UDI-Public(01)10705034773339(10)A7NA31
Combination Product (Y/N)N
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 03/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number388.509
Device Catalogue Number388.509
Device Lot NumberA7NA31
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2018
Initial Date Received by Manufacturer 03/15/2018
Supplement Date Received by Manufacturer04/17/2018
Initial Report FDA Received Date04/11/2018
Supplement Report FDA Received Date05/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-