| Model Number |
1098909 |
| Medical Device Problem Code |
Fire (1245)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/06/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned to date, however, photos of the damaged device battery have been received from the device distributor.These clearly indicate that thermal damage has occurred to the battery, but do not offer any indication as to the cause.
|
| |
|
Event or Problem Description
|
|
The patient reported a damage claim to their insurance company that there was property damage.The patient states that the device was charging, and then exploded.There was a small fire which caused some smoke damage and some burn marks in the carpeting.Internal reference ra (b)(4).
|
| |
|