MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAB; PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR Back to Search Results
Model Number MMT-723LNAB
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 03/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
The customer reported via phone call the insulin pump had multiple motor error alarms.The customer¿s blood glucose level was 328 mg/dl at the time of the incident and declined high blood glucose troubleshooting.The customer stated that the drive support cap was normal and that the insulin pump was not exposed to high magnetic fields.Customer reported a motor position encoder error.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The insulin pump is being replaced and is expected to return for analysis.
 
Additional Manufacturer Narrative
Pump alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform the displacement test, rewind, basic occlusion test, occlusion test, prime test and excessive no delivery test or verify the motor position encoder error alarm due to motor error alarms.Pump received with minor scratched lcd window, cracked case at the display window corners, cracked lcd window, cracked belt clip slot, broken battery tube threads, stained end cap sticker, stained address/serial number label and cracked reservoir tube lip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAB
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7425940
Report Number3004209178-2018-64441
Device Sequence Number2053192
Product Code OYC
UDI-Device Identifier00643169478565
UDI-Public(01)00643169478565
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 06/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-723LNAB
Device Catalogue NumberMMT-723LNAB
Device Lot NumberA4723LNABJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2018
Initial Date Received by Manufacturer 03/27/2018
Supplement Date Received by Manufacturer06/01/2018
Initial Report FDA Received Date04/12/2018
Supplement Report FDA Received Date06/27/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age60 YR
Patient Weight250
-
-