| Catalog Number |
024103 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/20/2016
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr is being filed after the associated complaint was reviewed under remediation protocol capa (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
|
| |
|
Event or Problem Description
|
|
It was reported that the physicians used the round tip ureteral catheter to enter the pediatric cystoscope in order to check the ureteral orifice.The physician was unable to push the catheter forward as a result of advancement difficulty when entering the cystoscope.After several failed attempts, the physician removed the catheter and noted that it was bent and fractured.It was further noted that the catheter was bent backwards and partially severed at a sideport that was 3 cm (centimeters) from the distal tip.The physician expressed that the catheter was too soft and the material was not very good.This product problem reportedly occurred prior to patient contact and there was no impact to the patient or end user.
|
| |
|