| Model Number |
X060-0320 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The complainant reported a complaint from (b)(6).The complainant stated that an axle screwdriver (x060-0320) broke when tightening the locking screw, and that the broken piece was recovered without incident.The complainant also added that the procedure was completed with a secondary screwdriver and that the patient was unaffected.A functionality assessment could not be performed due to the damage to the distal tip of the screwdriver.This screwdriver is meant to be used with a torque limiting handle, which reduces the force applied to the distal tip of the driver.The observed damage suggests that over-torquing may have contributed to the malfunction.A visual assessment of the screwdriver shows signs of repeated use.A dhr review was performed for lot#60190 and there were no manufacturing anomalies identified, and the device met required specifications.
|
| |
|
Event or Problem Description
|
|
The complainant reported a complaint from (b)(6).The complainant stated that an axle screwdriver (x060-0320) broke when tightening the locking screw, and that the broken piece was recovered without incident.The complainant also added that the procedure was completed with a secondary screwdriver and that the patient was unaffected.
|
| |
|