• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK UROLOGICAL INC FILIFORM DOUBLE PIGTAIL URETERAL STENT SET; FAD STENT, URETERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK UROLOGICAL INC FILIFORM DOUBLE PIGTAIL URETERAL STENT SET; FAD STENT, URETERAL Back to Search Results
Model Number G15120
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 04/12/2016
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This mdr is being filed after the associated complaint was reviewed under remediation protocol (b)(4), complaint/mdr retrospective review and remediation and reassessed as reportable.Additional complaint investigation and record remediation was not performed.
 
Event or Problem Description
The complainant reported that the pigtail tip of the filiform double pigtail ureteral stent was detached when the wire guide went through.The length of the detached stent was about 1cm (centimeter).The complaint device was replaced with a new device to finish the procedure.This product problem occurred prior to patient contact, and there was no impact to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FILIFORM DOUBLE PIGTAIL URETERAL STENT SET
Common Device Name
FAD STENT, URETERAL
Manufacturer (Section D)
COOK UROLOGICAL INC
1100 west morgan st
spencer IN 47460
MDR Report Key7426981
Report Number1820334-2018-01103
Device Sequence Number11181961
Product Code FAD
UDI-Device Identifier00827002151208
UDI-Public(01)00827002151208(17)181124(10)NS6406629
Combination Product (Y/N)N
Initial Reporter CountryCN
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other,use
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG15120
Device Catalogue Number133724
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2018
Initial Report FDA Received Date04/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-