• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. CHEMISTRY CALIBRATOR (CHEM CAL)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. CHEMISTRY CALIBRATOR (CHEM CAL) Back to Search Results
Model Number CHEMISTRY CALIBRATOR (CHEM CAL)
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A siemens headquarter support center (hsc) specialist reviewed the information in the complaint.There was not enough information provided to determine if the results were a comparison of calibrator lot 135808 to a previous calibrator lot or if the results were a comparison between two calibrations of the same calibrator lot.There was not enough information provided to determine if the bias or change in recovery is due to change in diagnosis, patient condition, change in calibrator lot, reagent lot or instrument performance.The cause of the discordant, falsely depressed creatinine result is unknown.Advia chemistry calibrator lot 135808 is performing as intended.No further evaluation of the device is required.Mdr 2432235-2018-00152 was filed for the same issue.
 
Event or Problem Description
A discordant, falsely depressed creatinine (enzymatic method) result was obtained on the advia chemistry enzymatic creatinine_2 (ecre_2) assay.The discordant result was obtained with a new lot of advia chemistry calibrator.The discordant result was not in alignment with the previous result from this sample.It is not known if the discordant result was reported to physician(s).There are no known reports of patient intervention due to the discordant creatinine result.There are no known reports of a delay in administering treatment or medical intervention to the patient due to the discordant creatinine result.
 
Additional Manufacturer Narrative
The initial mdr 2432235-2018-00151 was filed on 12-apr-2018.Additional information (28-may-2018): additional information was obtained from the siemens customer care center (ccc) specialist.The results obtained with advia chemistry calibrator lot 135808 were part of a comparison study between the results obtained with calibrator lot 135808 and the previous calibrator lot.Quality control materials were in range during the time of the comparison study.The cause of the discordant, falsely depressed creatinine result is unknown.Advia chemistry calibrator lot 135808 is performing as intended.No further evaluation of the device is required.Mdr 2432235-2018-00152_s1 was filed for the same issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHEMISTRY CALIBRATOR (CHEM CAL)
Common Device Name
CHEMISTRY CALIBRATOR (CHEM CAL)
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY 10591
MDR Report Key7426998
Report Number2432235-2018-00151
Device Sequence Number16542089
Product Code JIX
UDI-Device Identifier00630414223414
UDI-Public00630414223414
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K050374
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 06/01/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2019
Device Model NumberCHEMISTRY CALIBRATOR (CHEM CAL)
Device Lot Number135808
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer05/28/2018
Initial Report FDA Received Date04/12/2018
Supplement Report FDA Received Date06/01/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-