| Model Number |
MICL12.1 |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Codes |
Cataract, Induced (1767); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No similar complaint type events were reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that the surgeon implanted a 12.1 mm micl12.1, -16.0 diopter, implantable collamer lens into the patient's eye on (b)(6) 2017.The reporter stated that the surgeon will be removing the lens because it is too flat and it is causing a cataract.A longer lens in the same power was ordered to replace the lens.Additional information has been requested but none has been forthcoming.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
The surgeon has decided not to implant a larger lens since "is likely to cause narrow angles" and will monitor and consider cataract surgery as needed in the future.(b)(4).
|
| |
|