| Model Number |
MMT-523LNAL |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/25/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 182 at the time of the incident.The customer stated that the drive support cap appears sticking out.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform rewind test, basic occlusion, occlusion, prime/compromised force sensor system, excessive no delivery and displacement test or verify compromised force sensor system alarm due to completely broken reservoir tube lip.Also, the device had loose/protruded drive support disk and missing reservoir tube o-ring.
|
| |
|