Model Number MMT-523RNAH
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/22/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A test reservoir was installed and did not lock in place due to completely broken reservoir tube lip.Unable to perform the displacement test due to completely broken reservoir tube lip.Unit had minor scratched lcd window, cracked lcd window, cracked case at display window corner, cracked battery tube threads, missing reservoir tube o-ring, cracked reservoir tube, stained address/serial number label and missing end cap sticker.
 
Event or Problem Description
The customer reported via phone call that there was crack on the at the upper right screen area and reservoir lip was cracked and fixed to break off.The customer¿s blood glucose level was unknown.The insulin pump will be returned for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523RNAH
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key7428131
Report Number3004209178-2018-64883
Device Sequence Number112599
Product Code OYC
UDI-Device Identifier00643169476479
UDI-Public(01)00643169476479
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Patient
Type of Report Initial
Report Date (Section B) 04/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523RNAH
Device Catalogue NumberMMT-523RNAH
Device Lot NumberA0523RNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/30/2018
Initial Report FDA Received Date04/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age70 YR
Patient Weight137
-
-