| Model Number |
ADVIA CENTAUR XP |
| Medical Device Problem Codes |
Mechanical Problem (1384); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The cause of the discordant, falsely low thcg result on one patient sample is unknown.Siemens is investigating the issue.
|
| |
|
Event or Problem Description
|
|
A discordant, falsely low total human chorionic gonadotropin (thcg) result was obtained on one patient sample on an advia centaur xp instrument.The discordant result was reported to the physician(s), who questioned it.The patient sample was tested in another laboratory, which gave positive result for human chorionic gonadotropin.The customer repeated the sample on the same instrument and alternate advia centaur xp instrument, resulting higher.The correct results were reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely low thcg results.
|
| |
|
Additional Manufacturer Narrative
|
|
The initial mdr 2432235-2018-00150 was filed on 13-apr-2018.Additional information (27-mar-2018): a siemens customer service engineer (cse) was dispatched to the customer site.The cse performed a system check, verified aspiration and checked aspiration probes.The cse inspected the sample probe and adjusted the sample probe transducer voltage.The cse also checked the reagent and ancillary probes and performed other diagnostic checks and quality controls, which were acceptable.A siemens headquarters support center specialist reviewed the service report and concluded that the potential causes of the discordant result could have been no sample delivery or bleach contamination.The cse did not identify a system issue which would cause a falsely low result.The cause of the discordant, falsely low total human chorionic gonadotropin result on one patient sample is unknown.The instrument is performing within manufacturing specifications.No further evaluation of device is required.Section h6 has been updated with device, method, result, and conclusion codes.
|
| |
|
Search Alerts/Recalls
|