WRIGHTS LANE SYNTHES USA PRODUCTS LLC 6.0MM TI PRECONTOURED ROD 80MM STRAIGHT LENGTH/400MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DDD
|
Back to Search Results |
|
| Model Number |
04.600.018 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of postoperative rod breakage is unknown.Additional device product codes: mni, mnh, kwp, kwq.Implant date: implanted in 2014, exact date is unknown.(b)(4).The subject device has been received and the product evaluation is in progress.No conclusion can be drawn.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that a revision surgery was performed on a patient on (b)(6) 2018.The patient had a previous posterior t10-ilium surgery in 2014 with synthes uss (universal spine system) hardware.At an unknown point in time, the patient's right rod broke between the s1 and iliac screws.The left rod was not broken.During the revision, the surgeon removed the nuts and collars from the l4 and iliac screws bilaterally.He then cut the rods just caudal to the l5 screws.The surgeon also removed the left and right s1 screws.There was no allegation of deficiency or malfunction against the nuts and collars (4 each), or the s1 screws (2 each).Reported concomitant devices: uss nuts (part # 498.003, lot # unknown, qty.4); uss collars (part # 498.010, lot # unknown, qty.4); and uss screws (part # unknown, lot # unknown, qty.2).This report addresses revision surgery due to postoperative rod breakage.The intraoperative malfunctions which occurred during the revision surgery has been captured under linked complaint (b)(4).This is report 1 of 1 for complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A product investigation was conducted.Device condition: a piece of a precontoured titanium rod (part 04.600.018, lot unk, mfg unk) was received at us cq.The break is roughly transverse, approximately 1mm from the part number etch.Majority of the rod was not returned.The complaint condition of broken was visually confirmed and consistent with the reported condition.Replication of the complaint condition is not applicable as the devices were returned damaged.Lot number review: a dhr review could not be performed as the lot number is unknown.A material check was not performed at cq as there is no reason to think material propertied played a role since the device was implanted four years prior to breakage and explanation.Drawing review: relevant drawing(s) was reviewed.Dimensional analysis was completed, the outer diameter rod measured and is within specification based on relevant drawing.During the investigation no product design issues or discrepancies were observed that may have contributed to the complaint condition.Conclusion: the complaint condition is confirmed as the rod was received broken.A device inspection, drawing review and dimensional analysis were performed as part of this investigation.There is no indication that a design or manufacturing issue contributed to the complaint.While no definitive root cause could be determined it is possible that the device encountered unintended forces (such as trauma).No new malfunctions were observed during the course of this investigation.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|