| Model Number |
VLOCM0123 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, while doing continuous stitching, the device broke.There was no patient outcome.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned samples noted one suture and one needle.The needle was observed to be attached to the suture.The product was received with the loop open.It was observed, under magnification, that the loop appeared to have split under tension.Upon microscopic inspection, evidence of material transfer was found at the weld site of the loop.As no sealed products were returned, functional testing was precluded.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Should new information become available, the file will be re-opened, and the investigation summary amended as appropriate.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intra-operatively, while doing continuous stitching, the device broke.Another device was used to complete the case.There was no patient outcome.
|
| |
|