Model Number 482400S
Device Problem Incorrect Or Inadequate Test Results (2456)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer states that the enteral feeding pump failed the motor speed test.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed and the customer states, ¿fails motor speed test¿.The unit was triaged and the customer¿s reported condition was confirmed.The service technician found the pump had a defective sensor.A trend has been identified and a corrective action has been opened to address this issue.A review of the device history record shows that this unit was manufactured in 2011 and was released meeting all manufacturing specifications.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
789 puxing rd
building 10
shanghai 20111 4
CN  201114
Manufacturer (Section G)
COVIDIEN
789 puxing rd
building 10
shanghai 20111 4
CN   201114
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key7428944
MDR Text Key105520621
Report Number3006451981-2018-00273
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number482400S
Device Catalogue Number482400S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received03/22/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-