Model Number H7493926028400
Medical Device Problem Code Bent (1059)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device is a combination product.(b)(4).
 
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in a coronary artery.A 4.00 x 28mm synergy ii drug-eluting stent was advanced but failed to cross the lesion.When the device was removed from the patient's body, it was noted that the stent struts were flared.The procedure was completed with another 4.00 x 28mm synergy stent.No patient complications were reported and the patient's status was fine.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: synergy ii us, mr, 4.0x28mm, stent delivery system was returned for analysis.A visual examination of the crimped stent found stent strut rows 1 to 13 from the proximal end of the stent were damaged; some lifted and bent in a distal direction.The undamaged stent outer diameter is within the maximum crimped stent specification.A visual examination of the balloon cones was performed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination found multiple kinks along the hypotube shaft.This type of damage is consistent with excessive force being applied to the delivery system.A visual and tactile examination found no issue with the shaft polymer extrusion.A visual and tactile examination found no damage to the tip.No other issues noted during analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design & manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
It was reported that stent damage occurred.The target lesion was located in a coronary artery.A 4.00 x 28mm synergy ii drug-eluting stent was advanced but failed to cross the lesion.When the device was removed from the patient's body, it was noted that the stent struts were flared.The procedure was completed with another 4.00 x 28mm synergy stent.No patient complications were reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7429373
Report Number2134265-2018-03052
Device Sequence Number1718880
Product Code NIQ
UDI-Device Identifier08714729840312
UDI-Public08714729840312
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/23/2018
Device Model NumberH7493926028400
Device Catalogue Number39260-2840
Device Lot Number0020472268
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/02/2018
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer04/16/2018
Initial Report FDA Received Date04/13/2018
Supplement Report FDA Received Date05/07/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No