Model Number 04.501.003
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.Patient weight is unavailable for reporting.It is unknown when pain began.Implant date unknown.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent removal of plates and screws on (b)(6) 2018 due to pain.The patient requested removal of the three plates, screws, and five intermedullary splints as the patient was experiencing pain in the area of the implants and the patient believed the implants were the cause.The surgeon was able to remove the three plates that had been affixed with 22 screws; however, the five intermedullary splints and the five screws associated with the splints were not able to be removed.There was no surgical delay.The patient was stable after the surgery.This is report 1 of 6 for (b)(4).
 
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.A device history record (dhr) review was performed for part no.: 04.501.003, lot no.: 9537801, manufacturing location: (b)(4), release to warehouse date: 22.Jun.2015: no non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.A product investigation was performed.The ti matrixrib pre-contoured plate 16 holes for left ribs 4&5 (04.501.003) was received cut down to 12 holes and shows minor wear consistent with implantation and explantation.No damage was observed which may have inhibited intended device functionality.As no device defects or deficiencies were alleged it is not possible to confirm the complaint or replicate the complaint condition.Additionally, based on the allegation a drawing and dimensional analysis are not applicable as there is no indication that the device¿s design or manufacturing contributed to the reported complaint condition.A device history review, including material and hardness reviews, was performed for the returned implant¿s lot number and no mrrs, ncrs or complaint-related issues were identified with the lot number which may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI MATRIXRIB PRE-CONTOURED PL 16 HOLES FOR LEFT RIBS 4 & 5
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7429556
Report Number2939274-2018-51664
Device Sequence Number1124796
Product Code HRS
UDI-Device Identifier10887587016962
UDI-Public(01)10887587016962(10)9537801
Combination Product (Y/N)N
Initial Reporter StateGA
PMA/510(K) Number
K161590
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/22/2018
Device Model Number04.501.003
Device Catalogue Number04.501.003
Device Lot Number9537801
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2018
Initial Date Received by Manufacturer 04/13/2018
Supplement Date Received by Manufacturer04/13/2018
05/08/2018
Initial Report FDA Received Date04/13/2018
Supplement Report FDA Received Date04/13/2018
05/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
-
-