• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC HOOK OR SCREW HOLDER; MISC ORTHO SURGICAL INSTR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC HOOK OR SCREW HOLDER; MISC ORTHO SURGICAL INSTR Back to Search Results
Model Number 388.61
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 03/20/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient weight unknown device is an instrument and is not implanted/explanted.Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported during a revision surgery that the removal of s1 screws that two implant holders broke.Additionally, while replacing the right s1 screw, two universal spine system-variable angle screws (uss-vas) broke - (the locking ring on the screw became dislodged and would not work).A third implant holder broke during final tightening, and a tip of the holder broke off into the left s1 screw and was not retrievable.It remained in the patient.While inspecting the uss-vas screwdrivers, it was discovered that the tips of the drivers were twisted.Replacements were requested.The revision surgery was captured as a postoperative event on a related product complaint (b)(4).Concomitant devices reported: unknown uss screws (part unknown, lot unknown, quantity 3).This is report 3 of 7 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOOK OR SCREW HOLDER
Common Device Name
MISC ORTHO SURGICAL INSTR
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7429705
Report Number2939274-2018-51668
Device Sequence Number1768655
Product Code LXH
UDI-Device Identifier10705034773612
UDI-Public(01)10705034773612
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial
Report Date (Section B) 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number388.61
Device Catalogue Number388.61
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/20/2018
Initial Report FDA Received Date04/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-