• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. VACCESS PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. VACCESS PTA BALLOON DILATATION CATHETER Back to Search Results
Catalog Number VA8074
Medical Device Problem Codes Break (1069); Retraction Problem (1536); Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Patient Involvement (2645)
Date of Event 03/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No medical records and no medical images were provided to the manufacturer.The lot number of the device was provided.The device history records are currently under review.The device has been returned for evaluation.The investigation is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that during preparation for an angioplasty procedure, the balloon was allegedly found detached from the catheter upon removing the device from the packaging.There was no damage noted to the packaging.It was further reported that the procedure was performed with another device.There was no patient contact.
 
Additional Manufacturer Narrative
Manufacturing review: the lot number was provided and the lot device history records have been reviewed.The lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: one vaccess pta catheter was returned for evaluation.A visual inspection of the device found a circumferential break in the outer catheter shaft proximal the balloon.The inner guidewire lumen was noted to be intact.Therefore, the investigation is confirmed for the identified break and unconfirmed for the reported balloon detachment.The investigation is inconclusive for retraction issues as the 7fr introducer sheath was not returned for evaluation.The definitive root cause for the identified break could not be determined based upon available information.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.(removed patient code 2645, added patient code 2199).
 
Event or Problem Description
It was reported that during preparation for an angioplasty procedure, the balloon was allegedly found detached from the catheter upon removing the device from the packaging.There was no damage noted to the packaging.It was further reported that the procedure was performed with another device.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VACCESS PTA BALLOON DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key7429706
Report Number2020394-2018-00385
Device Sequence Number13966319
Product Code DQY
UDI-Device Identifier00801741121319
UDI-Public(01)00801741121319
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K111850
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/23/2021
Device Catalogue NumberVA8074
Device Lot Number93NC0120
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2018
Initial Date Received by Manufacturer 03/21/2018
Supplement Date Received by Manufacturer06/18/2018
Initial Report FDA Received Date04/13/2018
Supplement Report FDA Received Date06/25/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-