• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS PROV 4F DL IR 70CM NI; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS PROV 4F DL IR 70CM NI; CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS Back to Search Results
Catalog Number S3274355
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device has not been returned to the manufacturer for evaluation.A lot history review (lhr) of rebp1581 showed no other similar product complaint(s) from this lot number.(b)(4).
 
Event or Problem Description
Per medwatch, it was reported the patient's picc line was found to have a leak at the end of the red hub.The picc was removed and replaced.
 
Event or Problem Description
Per medwatch it was reported the patient's picc line was found to have a leak at the end of the red hub.The picc was removed and replaced.
 
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a leak is confirmed.One 4 fr dual lumen power picc provena was returned for evaluation.An initial visual observation showed the extension leg of the purple lumen appeared to have been cut approximately 2.4 cm proximal to the molded joint.Use residue was observed throughout the returned sample and a needleless injection cap and end cap were observed to be attached to the purple luer hub.A split was observed in the extension tubing of the red lumen just within the distal end of the luer connector.A microscopic observation revealed the fracture surface of the split in the extension tubing of the red lumen was rough and exhibited cracks/fissures and beach marks, which are indicative of material fatigue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROV 4F DL IR 70CM NI
Common Device Name
CATHETER,INTRAVASCULAR,THERAPEUTIC,LONG-TERM GREATER THAN 30 DAYS
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7429785
Report Number3006260740-2018-00698
Device Sequence Number10790328
Product Code LJS
UDI-Device Identifier00801741130052
UDI-Public(01)00801741130052
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K162443
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall,Other
Type of Report Initial,Followup
Report Date (Section B) 02/14/2019
Report Date (Section F) 03/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberS3274355
Device Lot NumberREBP1581
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2019
Event Location Hospital
Date Report to Manufacturer03/14/2018
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer02/06/2019
Initial Report FDA Received Date04/13/2018
Supplement Report FDA Received Date02/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age11 YR
-
-