• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS LIFELINE LIFELINE PERSONAL RESPONSE SYSTEM; SYSTEM, COMMUNICATION, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS LIFELINE LIFELINE PERSONAL RESPONSE SYSTEM; SYSTEM, COMMUNICATION, POWERED Back to Search Results
Model Number FD100
Health Effect - Clinical Codes Fall (1848); Bone Fracture(s) (1870); Hemorrhage/Blood Loss/Bleeding (1888); Loss of consciousness (2418)
Date of Event 03/13/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As reported to fda in 1220762-11-15-2017-001-r, the involved button was an affected device.The button did not have the appropriate fall parameters.It is unclear from the information received by the manufacturer if the button may have contributed to any injury the subscriber sustained.Per autoalert help button, instructions for use (p/n 0940817 rev.7 p.10 the autoalert help button can detect most falls, providing an added layer of protection.If you think you need assistance, push your help button immediately to initiate the help call pushing the help button generates the help call immediately.Recommended usage: push your autoalert help button any time you need help.If you fall and are able to, you should still push the autoalert help button to send a help call right away.Per autoalert help button, instructions for use, p/n 0940718 rev.5 page states: "in certain situations, the autoalert may not detect a fall.Some movements may not register as a fall and would not be detected." the involved button was still able to be depressed to summon help.
 
Event or Problem Description
The subscriber had a fall down several stairs.She was unconscious.She was bleeding from the fall and had broken several bones.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFELINE PERSONAL RESPONSE SYSTEM
Common Device Name
SYSTEM, COMMUNICATION, POWERED
Manufacturer (Section D)
PHILIPS LIFELINE
111 lawrence st
framingham MA 01702 8156
MDR Report Key7429876
Report Number1220762-2018-00001
Device Sequence Number13947290
Product Code ILQ
UDI-Device Identifier00884838084957
UDI-Public(01)00884838084957(11)141209(96)170925
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 03/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Patient Family Member or Friend
Device Model NumberFD100
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2018
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/19/2018
Initial Report FDA Received Date04/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-