• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.0MM HEXAGONAL SCREWDRIVER WITH T-HANDLE; SCREWDRIVERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 3.0MM HEXAGONAL SCREWDRIVER WITH T-HANDLE; SCREWDRIVERS Back to Search Results
Catalog Number 314.132
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient weight unknown.Device is an instrument and is not implanted/explanted.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported during a revision surgery that the removal of s1 screws that two implant holders broke.Additionally, while replacing the right s1 screw, two universal spine system-variable angle screws (uss-vas) broke - (the locking ring on the screw became dislodged and would not work).A third implant holder broke during final tightening, and a tip of the holder broke off into the left s1 screw and was not retrievable.It remained in the patient.While inspecting the uss-vas screwdrivers, it was discovered that the tips of the drivers were twisted.Replacements were requested.The revision surgery was captured as a postoperative event on a related product complaint (b)(4).Concomitant devices reported: unknown uss screws (part unknown, lot unknown, quantity 3).This is report 7 of 7 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Lot number dhr review was completed.Part no.: 314.132 , lot no.: l671242 , manufacturing location: haegendorf , release to warehouse date: 01.Dec.2017.No ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.Customer quality conducted an investigation of the returned device.It was reported that a 3.0mm hexagonal driver was found to be twisted.The 3.0mm hexagonal driver (314.132) is a component of the uss variable axis screw (vas) system for posterior fixation of the lumbar spine.The device is utilized for screw insertion.The returned instrument was examined and the complaint condition was able to be confirmed as the distal drive tip was found to be twisted.The failure mode of a twisted tip is likely the result of the application of excessive torque leading to deformation; hard patient bone or inadequate pedicle preparation may have contributed to the complaint conditions.The complaint condition cannot be replicated due to post-manufacturing damage.Relevant drawings for the returned device were reviewed (both current revision and from the time of manufacture): top-level.The design, materials and finishing processes were found to be appropriate for the intended use of this device.No dimensional analysis is possible due to post-manufacturing damage.A device history review, including material and hardness reviews, was performed for the returned instrument¿s lot number and no mrrs, ncrs or complaint-related issues were identified with the lot number which may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3.0MM HEXAGONAL SCREWDRIVER WITH T-HANDLE
Common Device Name
SCREWDRIVERS
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7429933
Report Number2939274-2018-51672
Device Sequence Number11204846
Product Code HXX
UDI-Device Identifier10705034769004
UDI-Public(01)10705034769004
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number314.132
Device Lot NumberL671242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2018
Initial Date Received by Manufacturer 03/20/2018
Supplement Date Received by Manufacturer05/08/2018
Initial Report FDA Received Date04/13/2018
Supplement Report FDA Received Date05/16/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age69 YR
-
-