• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC TI MATRIXRIB UNIVERSAL PLATE 8 HOLES; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC TI MATRIXRIB UNIVERSAL PLATE 8 HOLES; PLATE, FIXATION, BONE Back to Search Results
Model Number 04.501.009
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pain (1994)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device was used for treatment, not diagnosis.It is unknown when pain began.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient underwent removal of plates and screws on (b)(6) 2018 due to pain.The patient requested removal of the three plates, screws, and five intermedullary splints as the patient was experiencing pain in the area of the implants and the patient believed the implants were the cause.The surgeon was able to remove the three plates that had been affixed with 22 screws; however, the five intermedullary splints and the five screws associated with the splints were not able to be removed.There was no surgical delay.The patient was stable after the surgery.This is report 3 of 6 for (b)(4).
 
Additional Manufacturer Narrative
A device history record (dhr) review was performed for part no.: 04.501.009, lot no.: 9329055, manufacturing location: (b)(4), release to warehouse date: 21.Jan.2015: no non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.A product investigation was performed.The ti matrixrib universal plate 8 holes (04.501.009) shows minor wear consistent with implantation and explantation.No damage was observed which may have inhibited intended device functionality.As no device defects or deficiencies were alleged it is not possible to confirm the complaint or replicate the complaint condition.Additionally, based on the allegation a drawing and dimensional analysis are not applicable as there is no indication that the device¿s design or manufacturing contributed to the reported complaint condition.A device history review, including material and hardness reviews, was performed for the returned implant¿s lot number and no mrrs, ncrs or complaint-related issues were identified with the lot number which may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI MATRIXRIB UNIVERSAL PLATE 8 HOLES
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7430136
Report Number2939274-2018-51676
Device Sequence Number1216897
Product Code HRS
UDI-Device Identifier10887587017082
UDI-Public(01)10887587017082(10)9329055
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 03/22/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number04.501.009
Device Catalogue Number04.501.009
Device Lot Number9329055
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2018
Initial Date Received by Manufacturer 03/22/2018
Supplement Date Received by Manufacturer05/08/2018
Initial Report FDA Received Date04/13/2018
Supplement Report FDA Received Date05/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
-
-