WRIGHTS LANE SYNTHES USA PRODUCTS LLC TI MATRIXRIB UNIVERSAL PLATE 8 HOLES; PLATE, FIXATION, BONE
|
Back to Search Results |
|
| Model Number |
04.501.009 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pain (1994)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.It is unknown when pain began.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient underwent removal of plates and screws on (b)(6) 2018 due to pain.The patient requested removal of the three plates, screws, and five intermedullary splints as the patient was experiencing pain in the area of the implants and the patient believed the implants were the cause.The surgeon was able to remove the three plates that had been affixed with 22 screws; however, the five intermedullary splints and the five screws associated with the splints were not able to be removed.There was no surgical delay.The patient was stable after the surgery.This is report 3 of 6 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was performed for part no.: 04.501.009, lot no.: 9329055, manufacturing location: (b)(4), release to warehouse date: 21.Jan.2015: no non-conformance reports (ncrs) were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.A product investigation was performed.The ti matrixrib universal plate 8 holes (04.501.009) shows minor wear consistent with implantation and explantation.No damage was observed which may have inhibited intended device functionality.As no device defects or deficiencies were alleged it is not possible to confirm the complaint or replicate the complaint condition.Additionally, based on the allegation a drawing and dimensional analysis are not applicable as there is no indication that the device¿s design or manufacturing contributed to the reported complaint condition.A device history review, including material and hardness reviews, was performed for the returned implant¿s lot number and no mrrs, ncrs or complaint-related issues were identified with the lot number which may have contributed to the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|