U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC ADVANCE 14 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC ADVANCE 14 LP LOW PROFILE BALLOON CATHETER; DQY CATHETER, PERCUTANEOUS Back to Search Results
Catalog Number PTAX4-14-170-2.5-12
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/28/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, or unavailable.(b)(4).This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported that during an endovascular therapy procedure via the femoral approach, the balloon of the advance 14 lp low profile balloon catheter ruptured.Per the initial reporter, when the balloon reached the severely calcified lesion, it was inflated to 10 atmospheres (atm).Approximately 100 seconds after the start of inflation the rupture occurred.It is not known if the balloon ruptured longitudinally or circumferentially.Contrast used was visipaque in a 1:1 ration with saline.The procedure was successfully completed with another balloon device.The patient did not require any additional procedures due to this occurrence.No adverse events were reported.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation: a review of the dimensional verification, complaint history, device history record, drawing, instructions for use (ifu), quality control, specifications, and a visual inspection and functional evaluation of the returned device were conducted during the investigation.The visual inspection of the returned device confirmed the presence of a pinhole, which was noted in the balloon a distance of 11.4 cm from the distal end of the catheter.No other non-conformities were noted during the evaluation.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.It should be noted there were no other reported complaints for this lot number.Based on the information provided, the examination of the returned product, and the results of our investigation, a definitive root cause could not be determined.We will continue to monitor for similar complaints.Per the quality engineering risk assessment, no further action is required.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.Refer to additional mfr narrative for investigational findings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCE 14 LP LOW PROFILE BALLOON CATHETER
Common Device Name
DQY CATHETER, PERCUTANEOUS
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key7430299
Report Number1820334-2018-00733
Device Sequence Number2156207
Product Code DQY
UDI-Device Identifier10827002503240
UDI-Public(01)10827002503240(17)190329(10)6849703
Combination Product (Y/N)N
PMA/510(K) Number
K090822
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberPTAX4-14-170-2.5-12
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 03/28/2018
Supplement Date Received by Manufacturer07/18/2018
Initial Report FDA Received Date04/13/2018
Supplement Report FDA Received Date07/26/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-