AV-TEMECULA-CT XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT SYSTEM Back to Search Results
Catalog Number 1125300-28
Health Effect - Clinical Codes Angina (1710); Thrombosis (2100)
Date of Event 03/19/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.The reported patient effects of angina and thrombosis are listed in the xience alpine, everolimus eluting coronary stent system, instructions for use as known patient effects of coronary stenting procedures.The investigation determined the reported difficulty and patient effects, appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion in the mildly tortuous, mildly calcified, 75% stenosed, mid left anterior descending (lad) coronary artery using a 3.00 x 28 mm xience alpine stent delivery system (sds).The sds was advanced to the lesion and the stent implant was successfully deployed.Approximately one hour post procedure, the patient developed chest pain and a thrombus was confirmed by angiography and intravascular ultrasound (ivus) to be within the stent implant.There was a possibility that the apposition of the stent implant was not complete as it was an angiography guided percutaneous coronary intervention (pci).Thrombus aspiration was performed and an unspecified stent implant was deployed overlapping the proximal end of the xience alpine stent.The timi flow increased from 1 to 3 following the intervention.It was the physician's opinion that the relationship between the thrombus and the xience alpine stent was low.No additional information provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE ALPINE EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7430440
Report Number2024168-2018-02778
Device Sequence Number11250157
Product Code NIQ
UDI-Device Identifier08717648199370
UDI-Public08717648199370
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 04/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/06/2020
Device Catalogue Number1125300-28
Device Lot Number7102361
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2018
Initial Report FDA Received Date04/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-