(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was not returned for analysis.A review of the lot history records revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lots revealed no other incidents.The investigation determined the reported difficulties were likely due to case circumstances.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|
|
It was reported that the armada 018 balloon dilatation catheter became stuck on the 018 steel core guide wire during advancement to the uncalcified, non-tortous popliteal artery.The armada 018 was unable to be removed from the guide wire, so both were removed from the anatomy as a unit.Outside the anatomy, the devices were removed from each other.The same guide wire was used with a new armada 018 to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.There was no additional information provided.
|