• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEODENT - JJGC S.A. TI DRIVE ACQUA IMPLANT (3.3) 3.5X10; ENDOSSEOUS DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEODENT - JJGC S.A. TI DRIVE ACQUA IMPLANT (3.3) 3.5X10; ENDOSSEOUS DENTAL IMPLANT Back to Search Results
Catalog Number 140.753
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Tissue Damage (2104)
Date of Event 02/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Exemption number: e2015015.After the evaluation was noticed that the item received is different from the item reported on guarantee form by the dentist.Therefore, the lot number was not considered.Considering the surgical methodology, it was seen that the dentist has selected an implant design which is not recommended for the patient's bone quality.
 
Event or Problem Description
Rp 011125 - the dentist reported that 5 months after dental implant was installed in intraoral region 7#, its non-osseointegration was observed.In addition, the patient presented bone type ii.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI DRIVE ACQUA IMPLANT (3.3) 3.5X10
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key7430745
Report Number3008261720-2018-01706
Device Sequence Number1124809
Product Code DZE
UDI-Device Identifier07899878017080
UDI-Public07899878017080
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K133592
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2018
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 04/13/2018
Report Date (Section F) 03/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Dentist
Device Catalogue Number140.753
Other Device ID Number07899878017080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2018
Was the Report Sent to FDA? No
Distributor Facility Aware Date04/13/2018
Event Location Other
Date Report to Manufacturer03/20/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/20/2018
Initial Report FDA Received Date04/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
-
-