MAUDE Adverse Event Report: AV-TEMECULA-CT OMNILINK ELITE PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT OMNILINK ELITE PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number 1012624-19
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Internal file number - (b)(4): during processing of this complaint, attempts were made to obtain complete event, patient and device information.Incorrect anatomy.The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other incidents reported for this lot.The omnilink elite vascular balloon expandable stent system electronic instructions for use (ifu) states: the system is indicated for the treatment of atherosclerotic iliac artery lesions.The violation of the ifu does not appear to have contributed to the reported stent movement.The investigation was unable to determine a conclusive cause for the reported unstable stent implant.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that the patient underwent a procedure to treat a target lesion in the moderately calcified left subclavian artery.The 8.0 x 19 mm omnilink elite vascular stent delivery system was advanced; however, during positioning, the stent started to come off the delivery system balloon.As the stent was still partially on the delivery system balloon, the physician was able to pull the delivery system back a bit and get the stent mostly centered back on the balloon.The stent delivery system balloon was then inflated to 1.5 atmospheres (atm) to secure the stent to the balloon.The stent delivery system was then advanced to the target lesion and deployed at the desired location.Post procedure, the patient was doing well.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key7430801
Report Number2024168-2018-02786
Device Sequence Number2156899
Product Code NIO
UDI-Device Identifier08717648179280
UDI-Public08717648179280
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 04/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2020
Device Catalogue Number1012624-19
Device Lot Number7060341
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/22/2018
Initial Report FDA Received Date04/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age66 YR
-
-