• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION ENVEO R DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION ENVEO R DELIVERY SYSTEM; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number ENVEOR-N-US
Medical Device Problem Codes Material Separation (1562); Positioning Problem (3009)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/28/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product analysis: the delivery catheter system (dcs) was received with the capsule separated and closed.The handle was intact.When the deployment knob was deployed, the outer member moved away from the capsule.The capsule was completely separated at the proximal end of the capsule.Several voids are observed over the nitinol reinforcing frame all along the distal end to the proximal end of the capsule.The device was returned with the end cap/screw gear snap fit connected.Conclusion: the lot history record was reviewed and showed that this product met all manufacturing specifications for product released for distribution.No issues were identified that would have impacted this event.Capsule separation typically occurs due to excessive compressive forces applied during the valve loading process.This excessive compressive force is only experienced when the valve is loaded incorrectly.However, the quality of load cannot be assessed as no fluoroscopic load check images were submitted for review, therefore it cannot be independently determined if the valve was correctly loaded onto the delivery catheter system (dcs).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information that during the implant of a transcatheter bioprosthetic valve, the valve was recaptured twice due to improper depth.Upon the third attempt at deployment, the delivery catheter system (dcs) would not unsheath.It was reported that it may have gone past the point of no recapture.The capsule was not moving when the handle was moving.The capsule was fully closed and the delivery catheter system (dcs) was removed from the patient.It was reported that the capsule was split in half with the valve stuck in the capsule.The capsule was broken near the proximal portion of the capsule, however no pieces remained in the patient.No resistance was felt by the physician and it was reported the patient had a tortuous and calcified anatomy.A different valve and delivery catheter system (dcs) were used for a successful deployment.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENVEO R DELIVERY SYSTEM
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key7430886
Report Number2025587-2018-00865
Device Sequence Number2156938
Product Code NPT
UDI-Device Identifier00643169640337
UDI-Public00643169640337
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
P130021
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/20/2018
Device Model NumberENVEOR-N-US
Device Catalogue NumberENVEOR-N-US
Device Lot Number0008459339
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/15/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/23/2018
Initial Report FDA Received Date04/13/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age67 YR
-
-