| Catalog Number |
364979 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/31/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Results - bd received samples and photos from the customer facility for investigation.The samples and photos were evaluated and the customer¿s indicated failure mode for foreign matter with the incident lot was observed.Additionally, a review of the manufacturing records was completed for the incident lot and no issues were identified.Conclusion - based on the investigation, a root cause could not be determined within the scope of this evaluation.The fm was identified as blood through pcr analysis by r&d; however, the source of the blood could not be traced back to any issues during manufacturing.
|
| |
|
Event or Problem Description
|
|
It was reported that bd vacutainer® urinalysis urine tube had foreign matter on the stopper.No injury or medical intervention reported.
|
| |
|