Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.If implanted, give date: unknown, information not provided.If explanted, give date: unknown, information not provided.(b)(4).Attempts have been made to obtain missing information; however, to date, no response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
|
Additional information: device available for evaluation ¿ yes, returned to manufacturer on 4/9/2018.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual the product was received stuck in cartridge.Visual inspection of the return sample shows the product is damaged, most probably due to the product was received stuck in cartridge.Considering the condition of the lens additional analysis is not possible.The customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints were received for this production order number.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|